How to find a suitable 'constant'
To arrive at a 'constant' which will make the measured numbers come closer to the calculated (hm^2/R) numbers, try the following:
In this example the hm^2/R values for the four zones were calculated to be as follows:
| zone 1 | zone 2 | zone 3 | zone 4 |
| 0.0025 | 0.0229 | 0.0590 | 0.1109 |
The numbers taken from the 4 zones of the mask used during testing gave the following result
| zone 1 | zone 2 | zone 3 | zone 4 |
| 0.152 | 0.172 | 0.209 | 0.260 |
Take the reading from the 3rd 'window' in the mask (zone 3 ) )which is usually the most accurate of the readings you will make. (If you have say 5 zones then take the 4th zone as being the most accurately read measurement.)
So you take the third reading number and subtract the hm^2 number calculated for the third window which let us say is 0.059 then subtract it from the reading number three(0.209) this gives you a 'constant' of 0.150 If you now subtract this same number 0.150 from all the readings you will get the following:
| zone 1 | zone 2 | zone 3 | zone 4 |
| 0.0020 | 0.0220 | 0.0590 | 0.1100 |
As long as you subtract the same 'constant' number from each of the readings - you will still maintain the difference of relative spacing between them, but they become easier to compare with the calculated numbers taken from the hm^2/R value for each window.
return to testing - calculations